The short version
- ISO 13485 asks you to document how competence is built, training is given and awareness is ensured, and to check effectiveness in proportion to the risk of the work.
- Training records must trace to the product: for one lot, which revision of the SOP the operators were trained on.
- One training matrix puts the role, the required courses, the retraining interval and the way of checking on one page.
1Where clause 6.2 differs from clause 7.2 of ISO 9001
Clause 6.2 of ISO 13485 also covers the competence of people, but it asks for more.
- A documented process for establishing competence, providing training and ensuring awareness.
- Methods of checking that training was effective, in proportion to the risk of the work.
- Records controlled under clause 4.2.5: identifiable, retrievable and kept for the required period.
Put simply, ISO 9001 asks whether you did it. ISO 13485 also asks you to write down how you do it, and to decide how deep to go by the risk.
2Match the check to the risk
The same course carries a different risk in a different job. A person who applies labels and a person who runs sterilization equipment should not be checked in the same way.
- Low risk jobs: a quiz after the course, plus a supervisor's confirmation.
- Medium risk jobs: a quiz, plus a practical assessment on the job.
- High risk jobs, for example processes that affect sterility or product function: a practical assessment and an independent check, and no solo work until the person is qualified.
The basis for the levels belongs in the procedure and should agree with the outcome of risk management. An auditor will ask why this job only has a written test.
3Records must trace to the product
A medical device audit often starts from one lot number. The auditor asks: when this lot was made, which revision of the SOP had the operators been trained on?
To answer, the training record needs the person, the SOP revision and the date, and it must line up with the dates of the production record. If weak training leads to a complaint or a corrective and preventive action (CAPA), the record should link to that case and show who else needs retraining.
Retention follows clause 4.2.5: at least the lifetime of the medical device, and not less than two years from release of the product, or as the applicable regulation requires.
4The US FDA QMSR and ISO 13485
The US FDA Quality Management System Regulation (QMSR) took effect on 2 February 2026. It incorporates ISO 13485:2016 by reference and replaces the earlier QSR text. For makers who export to the US, this means the personnel and record requirements of ISO 13485 become the yardstick at an inspection.
For details and scope, rely on the text of the regulation and the agency's own guidance.
5Taiwan medical device quality rules
Taiwan medical device quality rules are based on ISO 13485. For the clause mapping and the scope, rely on the latest version issued by the competent authority.
In practice, a maker builds one set of documents against two sets of requirements: the training procedure and the records follow ISO 13485, and at an inspection they are checked against the authority's text item by item.
6A sample training matrix
A training matrix puts the required courses, the retraining interval and the way of checking for each job on one page. The table below is an example. Adjust the jobs and intervals to your own procedure.
| Job | Required courses | Retraining interval | How it is checked |
|---|---|---|---|
| Assembly operator | Induction; assembly SOP | Yearly | Quiz and supervisor sign off |
| Quality inspector | Incoming inspection; use of measuring instruments | Yearly | Quiz and practical assessment |
| Cleanroom operator | Cleanroom conduct; gowning procedure | Yearly | Quiz and observation on the job |
| Process engineer | Process change control; risk management | When an SOP changes | Supervisor review |
Assembly operator
- Required courses
- Induction; assembly SOP
- Retraining interval
- Yearly
- How it is checked
- Quiz and supervisor sign off
Quality inspector
- Required courses
- Incoming inspection; use of measuring instruments
- Retraining interval
- Yearly
- How it is checked
- Quiz and practical assessment
Cleanroom operator
- Required courses
- Cleanroom conduct; gowning procedure
- Retraining interval
- Yearly
- How it is checked
- Quiz and observation on the job
Process engineer
- Required courses
- Process change control; risk management
- Retraining interval
- When an SOP changes
- How it is checked
- Supervisor review
The matrix lets an auditor see the rule at a glance, and lets you see who is missing what.
Remember this
ISO 9001 asks whether you trained. ISO 13485 also asks how you decided how far to train, and whether the record traces to that lot.